Siemens AG
SPR2-330.815.01.01.02
ARCADIS Avantic
02.05
CS SD 24
General Startup
17
Page 17 of 46
Medical Solutions
Function check of the EMERGENCY STOP switches
0
•
Connect the system to the voltage and boot it.
•
Click "Emergency" for the patient registration.
•
Activate the EMERGENCY STOP switch on the C-arm.
•
The safety switch moves audibly and the following message appears on the lower left
monitor edge: "The emergency stop has been pressed"
•
Disengage the EMERGENCY STOP switch.
•
The safety switch moves audibly and the following message appears on the lower left
monitor edge: "The emergency stop switch has been released."
•
Document the function of the EMERGENCY STOP switch in the "Test protocol, C-arm
movement and emergency stop"
(C-arm movement and emergency stop / p. 38)
.