16
General Startup
ARCADIS Avantic
SPR2-330.815.01.01.02
Siemens AG
02.05
CS SD 24
Page 16 of 46
Medical Solutions
Fig. 3:
Measuring circuit for the direct measurement of the system leakage current
•
Measurement configuration according to
application part
is not available in ARCADIS Avantic.
•
For safety reasons, place the system on an insulated base.
•
Perform measurement and repeat measurements.
•
Document the measured values in the "Leakage current test protocol",
.