General Startup

13

Page 13 of 46

Medical Solutions

3-

 

2General Startup

Power Connection

0

Measuring the Power Line Voltage

0

Measure the voltage and frequency at the intended outlet for the ARCADIS Avantic sys-
tem.

Compare the measured line voltage with the line voltage label on the back of the moni-
tor cart.

¹

The measured voltage must agree with the voltage read off from the label.

Summary of Contents for ARCADIS Avantic SP

Page 1: ...authority Offenders will be liable for damages All rights including rights created by patent grant or registration of a utility model or design are reserved Print No W ü r t h w e i n ARCADIS Avantic System Startup SP 2005 SPR2 330 815 01 01 02 Replaces n a English Doc Gen Date 02 05 Start up Instructions 10047680 10048590 ...

Page 2: ...to order current revision levels Disclaimer The installation and service of equipment described herein is to be performed by qualified personnel who are employed by Siemens or one of its affiliates or who are otherwise authorized by Siemens or one of its affiliates to provide such services Assemblers and other persons who are not employed by or otherwise directly affiliated with or authorized by S...

Page 3: ...line voltage can be adapted at the transformer in the monitor cart 14 Adapting the line voltage 14 Measuring the system leakage current 15 Measuring arrangements for system leakage current with tester VDE 0751 October 2001 preferred measuring method 15 Measuring arrangements for system leakage current with tester VDE 0751 October 1990 e g Bender tester 15 Function check of the EMERGENCY STOP switc...

Page 4: ...leting protocols 37 Completing the ARCADIS Avantic Installation protocol 37 5 _______ Protocols _____________________________________________________ 38 C arm movement and emergency stop 38 Leakage current measurement 39 Network checklist 41 6 _______ Changes to Previous Version _____________________________________ 45 ...

Page 5: ...Information 5 Page 5 of 46 Medical Solutions 2 1General Information Additionally Required Documents 0 ARCADIS Avantic Operator Manual ARCADIS Avantic Wiring Diagram IQ test quality assurance Image intensifier laser light localizer optional assembly and setting instructions ...

Page 6: ...st kit 37 90 156 X1963 Copper filter set 44 06 120 RV090 Densitometer 97 02 416 Y1996 Radiation filter set incl 25 mm Al 97 98 596 G5321 Resolution test 28 71 820 RE999 Safety tester leakage current and ground wire testing device e g Unimed 1100 Leakage current measurement only required in the case of a deviation of the local line voltage from the system delivery sta tus 51 38 727 Y0766 ...

Page 7: ...DIS Avantic 02 05 CS SD 24 General Information 7 Page 7 of 46 Medical Solutions Note on the log book 0 The log book is located in the monitor cart behind the keyboard It is accessible if the cover on the back of the monitor cart is removed ...

Page 8: ...e operation of the ARCADIS Avantic refer to current and voltage values stated on the line voltage label back of monitor cart The required line internal resistance is reached refer to ARCADIS Avantic Planning Guide The selectivity of the fuse protection is assured according to the relevant national standards It must also be ensured by the customer that connection to an unsuitable power supply is pr...

Page 9: ...k of the monitor cart Necessary adjustments to local conditions are to be performed as required according to the adjustment instructions of the ARCADIS Avantic system WARNING Danger high voltage Disregarding safety precautions can result in death or serious bodily injury Before the line voltage is adapted the power plug on the UPS in the monitor cart must be pulled out since the monitor cart is no...

Page 10: ...are to be labeled invalid Check the comment values invalid and confirm with name date and signature The equivalent leakage current measurement must be repeated Perform the measurement according to DIN VDE 0751 part 1 see ARTD part 2 and record the determined value as the initial measured value This value must not exceed the maximum value of 2 mA for equivalent leakage currents for devices accordin...

Page 11: ...ed period afterward voltage is still present in the electrolyte condenser battery and the generator boards D20 D21 and D30 Before work is started on the electrolyte condenser bat tery it must must be discharged To do this switch D21 S2 must be set to the UZ_OFF position LEDs D21 X22 and D21 X23 must not continue to glow Then the voltage between points X109 UZ_IST ACT and X109 ANA_GND is measured T...

Page 12: ...s for Techni cal Service X These checks and adjustments are explicitly designated on the following pages with the radiation warning sym bol CAUTION Laser emissions This product contains two class 2 lasers USA Laser class 2 as well as class 1 M image intensifier light localizer Disregarding safety precautions can lead to bodily injury espe cially to the retina of the eye resulting in irreversible d...

Page 13: ...eral Startup Power Connection 0 Measuring the Power Line Voltage 0 Measure the voltage and frequency at the intended outlet for the ARCADIS Avantic sys tem Compare the measured line voltage with the line voltage label on the back of the moni tor cart The measured voltage must agree with the voltage read off from the label ...

Page 14: ... the line voltage is adapted the power plug on the UPS in the monitor cart must be pulled out since the monitor cart is not free of voltage after the power plug is pulled out Adapting the line voltage 0 Remove the back bottom covers Fig 1 p 14 from the monitor cart Pull the power plug out of the UPS in the monitor cart Adapt the voltage by changing the connections to transformers T1 and T2 Refer t...

Page 15: ...ngements for system leakage current with tester VDE 0751 Octo ber 2001 preferred measuring method 0 Fig 2 Measuring circuit for measuring the system leakage current according to the differential current method Perform measurement configuration according to Fig 2 p 15 Application part 1 Fig 2 p 15 is not available in ARCADIS Avantic Perform measurement and repeat measurements Document the measured ...

Page 16: ...easurement of the system leakage current Measurement configuration according to Fig 2 p 15 application part 1 Fig 3 p 16 is not available in ARCADIS Avantic For safety reasons place the system on an insulated base Perform measurement and repeat measurements Document the measured values in the Leakage current test protocol Leakage cur rent measurement p 39 ...

Page 17: ...STOP switch on the C arm The safety switch moves audibly and the following message appears on the lower left monitor edge The emergency stop has been pressed Disengage the EMERGENCY STOP switch The safety switch moves audibly and the following message appears on the lower left monitor edge The emergency stop switch has been released Document the function of the EMERGENCY STOP switch in the Test pr...

Page 18: ...ns Functional checks 0 Operating the ARCADIS Avantic refer to the Operator Manual Stand and C arm system movements X Semi transparent slot and iris diaphragm in fluoroscopy procedure X Collimation of the semi transparent slot and iris diaphragm using fluoroscopy proce dure Mechanical functions of the monitor cart ...

Page 19: ...rs 0 The system must be free of voltage Measuring de vice Protective conductor testing device e g safety tester Unimed 1100 Measuring de vice Max 0 2 Ohm observe country specific regulations current 10 A Test procedure Measure between all touchable conductive parts of the system and the protective conductor wire at the ARCADIS Avantic power plug ...

Page 20: ... press the downward button once again The lifting column can then be moved by a further 7cm to the lowest point position 2 2 Fig 4 p 20 In this area a signal 3 beeps sounds for safety reasons each time the downward button is pressed To move the C arm upward press the upward button No signal sounds in this case When both switches are pressed the lifting motor moves in the initially selected direc t...

Page 21: ...m Settings 21 Page 21 of 46 Medical Solutions 4 3Specific System Settings System configuration 0 Please clarify in advance with the customer whether the following system configura tions are required Service login 0 Fig 5 Service login Select Options Customer service Local service ...

Page 22: ...ic System Settings ARCADIS Avantic SPR2 330 815 01 01 02 Siemens AG 02 05 CS SD 24 Page 22 of 46 Medical Solutions Fig 6 Authentication_ Enter the 6 character password and click OK see system folder for password ...

Page 23: ...Siemens AG SPR2 330 815 01 01 02 ARCADIS Avantic 02 05 CS SD 24 Specific System Settings 23 Page 23 of 46 Medical Solutions Customer address configuration 0 Fig 7 Configuration_ Select Configuration ...

Page 24: ...24 Specific System Settings ARCADIS Avantic SPR2 330 815 01 01 02 Siemens AG 02 05 CS SD 24 Page 24 of 46 Medical Solutions Fig 8 Configuration Click Next ...

Page 25: ...Siemens AG SPR2 330 815 01 01 02 ARCADIS Avantic 02 05 CS SD 24 Specific System Settings 25 Page 25 of 46 Medical Solutions Fig 9 Local Host Click Site Info ...

Page 26: ... 815 01 01 02 Siemens AG 02 05 CS SD 24 Page 26 of 46 Medical Solutions Fig 10 Site info The marked customer specific data may be adapted Do not change Serial No Click Save Click Home Main system configuration 0 In the service software select Main System ...

Page 27: ...R2 330 815 01 01 02 ARCADIS Avantic 02 05 CS SD 24 Specific System Settings 27 Page 27 of 46 Medical Solutions Fig 11 Main_system Click Main system Fig 12 Mainsys start Select ARCADIS Avantic I I type and click Next ...

Page 28: ...on time has elapsed Block time limit After the entered radiation time has elapsed without reset of the warning time limit the radiation is blocked the next time it is selected Every reset of Warning Time Limit also resets the Block Time Limit to 0 Buzzer mode indirect I I Buzzer off Buzzer sound off Exceptions Push mode and Warning Time Limit has ended Buzzer on Buzzer sound always during radiatio...

Page 29: ...t service program and log in as described under Service Login Select Main System Configuration Main System Footswitch CFC right Save the changes with Save Click Home If any reprogramming has been performed the label attached to the footswitch must be removed and replaced by the new label the replacement label is attached to the footswitch with a cable tie If a change is made to the pedal configura...

Page 30: ...ystem Settings ARCADIS Avantic SPR2 330 815 01 01 02 Siemens AG 02 05 CS SD 24 Page 30 of 46 Medical Solutions Fig 14 Main_system Click Main system Fig 15 Mainsys start Select ARCADIS Avantic I I type and click Next ...

Page 31: ...15 01 01 02 ARCADIS Avantic 02 05 CS SD 24 Specific System Settings 31 Page 31 of 46 Medical Solutions Fig 16 Current limitation Click Current limitation Make country specific changes only Denmark and click Save Click Home ...

Page 32: ...dical Solutions Backup 0 NOTE A backup must be created following software installation every software update and every system adjustment Select Backup Restore in the Service Software Home Menu Place the backup CD located in the monitor cart service compartment in the CD ROM drive Fig 17 Backup Restore_Command ...

Page 33: ...D 24 Specific System Settings 33 Page 33 of 46 Medical Solutions Select the Backup command Fig 18 Backup Restore_Command_Backup Select the respective drive under Drives Save to CD ROM standard save Fig 19 Backup SW Settings02 Select SW Settings02 under Packages ...

Page 34: ...ntil Ready is displayed in the footer Exam set Select ExamSet under Packages Click Go Wait until Ready is displayed in the footer Security settings For systems with HIPAA configuration Select Security Settings under Packages Click Go Wait until Ready is displayed in the footer Main system Select MainSystem under Packages Wait until Ready is displayed in the footer Exiting backup Close the window v...

Page 35: ...tic 02 05 CS SD 24 Specific System Settings 35 Page 35 of 46 Medical Solutions NOTE It is not possible to perform a backup of the database patient im ages Store the patient images in an archive Saving to CD ROM as a long term archive is prohibited ...

Page 36: ...he supplied printer was preinstalled at the factory Sony printer UP D72 X 0 Perform the following steps according to the user manual for the printer CD ROM Unpack the printer Fill the paper tray with paper follow instructions on paper package Plug in the USB cable Print a test page after the automatic test Follow the instructions on the CD ROM from the manufacturer in the event of problems ...

Page 37: ...ng all adjustment work steps and check measurements the responsible technician must sign and date the test protocols and country specific protocols to confirm that all values have been correctly determined and recorded Completing the ARCADIS Avantic Installation protocol 0 The manufacturer of this product requires information and the legislator demands proofs that a product delivered free of defec...

Page 38: ...arm movement and emergency stop 0 System ARCADIS Avantic Tab 1 C arm movement and emergency stop protocol Material no Serial number Movement OK Yes No Name Date Initials Releasing locking the brakes Downward movement position 1 Downward movement position 2 The warning signals sound as described Function EMERGENCY_STOP_button ...

Page 39: ... measurement Material no Serial number Equivalent leakage cur rent Measuring instrument used serial no Name Date Initials First measured value Repeat measurement 1 Repeat measurement 2 Repeat measurement 3 Repeat measurement 4 Repeat measurement 5 Repeat measurement 6 Repeat measurement 7 Repeat measurement 8 Repeat measurement 9 Repeat measurement 10 Repeat measurement 11 Repeat measurement 12 Re...

Page 40: ...D 24 Page 40 of 46 Medical Solutions Repeat measurement 14 Repeat measurement 15 Repeat measurement 16 Repeat measurement 17 Repeat measurement 18 Repeat measurement 19 Repeat measurement 20 Equivalent leakage cur rent Measuring instrument used serial no Name Date Initials ...

Page 41: ... network data Tab 3 System network data Network data for HIS RIS network partner only with HIS RIS option Network checklist 1 For SAP order no _______________________ Will the system have a network connection HIS RIS Camera PACS Yes No Name Phone fax E mail Not networked Identification Computer name Configuration Local Host TCP IP address IP address TCP IP address Configuration local host TCP IP a...

Page 42: ...Application entity AE title Configuration DICOM HIS RIS nodes Port number Configuration DICOM HIS RIS nodes Comment Information on each Reporting console Archive system Not available Product Version Manufacturer Host properties Host name Configuration DICOM network nodes TCP IP address Configuration DICOM network nodes General node properties Logical name Configuration DICOM network nodes Applicat...

Page 43: ...n Manufacturer Host properties Host name Configuration DICOM network nodes TCP IP address Configuration DICOM network nodes General node properties Logical name Configuration DICOM network nodes Application entity AE title Configuration DICOM network nodes Port number Configuration DICOM network nodes Supported Dicom Services Storage yes no Configuration DICOM network nodes Storage commitment yes ...

Page 44: ...uration DICOM Print devices Application entity AE title Configuration DICOM Print devices Port number Configuration DICOM Print devices Comment Information for Any camera No camera available Device Version Manufacturer Default camera yes no Host properties Host name Configuration DICOM Print devices TCP IP address Configuration DICOM Print devices General node properties Logical name Configuration...

Page 45: ...Siemens AG SPR2 330 815 01 01 02 ARCADIS Avantic 02 05 CS SD 24 Changes to Previous Version 45 Page 45 of 46 Medical Solutions 7 6Changes to Previous Version n a first version ...

Page 46: ...46 Changes to Previous Version ARCADIS Avantic SPR2 330 815 01 01 02 Siemens AG 02 05 CS SD 24 Page 46 of 46 Medical Solutions ...

Reviews: