Medical Solutions
Functional test of C-arm lift movement and parking brakes
0
Fig. 4:
Function test
•
Switch on the main system (C-arm).
•
To move the C-arm downward, press the downward button.
•
The lifting column moves to position 1
and remains there automatical-
ly. A stop signal (3 beeps) sounds simultaneously.
•
To lower the lifting column further, press the downward button once again.
•
The lifting column can then be moved by a further 7cm to the lowest point, position 2
. In this area a signal (3 beeps) sounds for safety reasons each time the
downward button is pressed.
•
To move the C-arm upward, press the upward button. No signal sounds in this case.
•
When both switches are pressed, the lifting motor moves in the initially selected direc-
tion.
•
Loosen the different parking brakes and apply them again. Secure blocking of the rele-
vant movement must be assured.
•
Document the function of the C-arm movements, brakes and warning signals in the test