18
8027092/AE00/V1-0/2022-10| SICK
O P E R A T I N G I N S T R U C T I O N S | VISIC100SF
Subject to change without notice
6
MENUS
6.7.3
Menu - Operating Data
Function: Read out operating data of the gas sensor
Display operating data
Procedure:
1
Menu: Installation & Calibration
→
Open Operating Data.
>>> Operating data are read in.
>>> “Please wait...” is displayed during reading in.
>>> Read-in not successful: Display returns to the previous menu item.
>>> Read-in successful: First subordinate menu item is displayed.
2
Press “Down arrow”to display the following menu items.
Symbol
Description
Software Version
Current software version
Days of Operation
Number of current operating days
Gas Conc. Counter
●
Gas concentration counter*
●
Unit. 10% of measuring range per day
Days of Operation expected
●
Expected lifetime of the gas sensor.
●
Is continuously recalculated based on adjustments and
concentration counter
Min. Temperature
●
Display of the lowest measured temperature (start value = 70°C)
●
Will be updated only after 7 days when restarting
Max. Temperature
●
Display of the highest temperature value measured on the device
(start value = -34°C)
●
Will be updated only after 7 days when restarting
Last Tool No.
Display of the tool number used for the last access.
Number of Calibr.
Number of adjustments performed
Sensitivity
Sensitivity of the sensor at the last adjustment to new sensor
Maintenance Days last
Remaining maintenance days due before the last adjustment
Days of operation Last
Number of operating days since last adjustment (fixed value)
Maintenance Days Currently Current remaining days until next maintenance date (down counter)
Max. Current Value
Maximum measured current value of the gas sensor