10
8027092/AE00/V1-0/2022-10| SICK
O P E R A T I N G I N S T R U C T I O N S | VISIC100SF
Subject to change without notice
3
PRODUCT DESCRIPTION
3.2
Layout and function
The Gas Adjustment Kit continuously checks the gas type and measuring range set for
agreement with gas type and measuring range of the connected gas sensor and outputs an
error if there is an inconsistency.
3.2.1
Gas Adjustment Kit
Fig. 2: Contents, Gas Adjustment Kit
Designation
1
Hose (marked blue/orange),
3 m for nitrogen monoxide (NO) and nitrogen dioxide (NO
2
)
2
Hose, transparent (marked pink),
3 m for carbon monoxide (CO) and synthetic air
3
Accessories: 2 × hose adapters and transition pieces,
2 × replacement seals (O-rings) for tapping valves
4
Stainless steel sampling valve (marked blue) for NO and NO
2
, flow rate: 0.5 l/min
5
Brass sampling valve (marked pink) for CO and synthetic air, flow rate: 0.5 l/min
6
Adjustment tool
Operating Instructions
Note:
It is assumed that the test gases are provided by the customer.
Calibration gas data,
.