8008250/1F7Y/V4-0/2022-06 | SICK
O P E R A T I N G I N S T R U C T I O N S | GM901
Subject to change without notice
MENUS
9
9.3.11
Changing the calibration values
Display
Action
Note
▸
Select “
Calibration
”
▸
Press “
Enter
”
On-site calibration
▸
Press “
Enter
”
SPAN / Characteristic Curve
Basic factory setting: 1.00
Change possible, e.g., after successful
Span test
▸
Select with “
Arrow right
”
▸
Enter new value with “
Arrow down
” or
“
Arrow up
”
▸
Press “
Enter
”
Confirm with “
Enter
” to save new value
Possible setting range for Span:
0.50 ... 1.99
▸
Press “
Enter
”
Basic factory setting: 0
Offset correction possible, e.g., after a
reference measurement
▸
Select with “
Arrow right
”
▸
Enter new value with “
Arrow down
” or
“
Arrow up
”
▸
Press “
Enter
”
Confirm with “
Enter
” to save new value
▸
Press “
Arrow left
” (back)
Display returns to selection screen
Parameters
back
Humidity
Pressure
Parameter Settings
back
Analog Out
Analog Out
Calibration
Parameters
back
Span : 1.00
Span : 1.00
Calibration
back
edit: Enter
edit: Enter
Zero :
0
Zero :
0
Parameters
back
Span : 1.00
Span : 1.00
Calibration
back
Zero :
0
Zero :
0
select
Parameters
back
Span : 1.00
Span : 1.00
Calibration
back
edit: Enter
edit: Enter
Zero :
0
Zero :
0
Parameters
back
Span : 1.00
Span : 1.00
Calibration
back
Zero :
0
Zero :
0
edit: Enter
edit: Enter
Parameters
back
Span : 1.00
Span : 1.00
Calibration
back
Zero : + 0 0 0
Zero : + 0 0 0
select
Parameters
back
Span : 1.00
Span : 1.00
Calibration
back
Zero :
0
Zero :
0
edit: Enter
edit: Enter