24
8008250/1F7Y/V4-0/2022-06 | SICK
O P E R A T I N G I N S T R U C T I O N S | GM901
Subject to change without notice
5
MOUNTING
5.4.1
Installing the standard flange
▸
Mark the assembly position of the “flange with tube”, cut a hole with a blowtorch.
▸
The supports for the “flange with tube” should protrude approx. 30 mm into the duct. If
necessary, adjust the tube supports.
▸
Tack-weld the flange with tube and maintain the exact measuring distance flange-flange
and dimensions for the “Top” marking of the fitting position.
Fig. 4: Flange with tube, standard version
▸
To align the flange with tube on-site: Use a tube (
) or the adjustment device.
Fig. 5: Aligning the flange using an optical adjustment device
NOTE:
Damage to the duct opening possible!
Make sure parts cut off do not fall into the duct
30 (1.2”)
∅
76
(1
.2”)
Marking
for fitting position
Position of threaded pins
shown offset 45°
Standard flange with tube
L [mm]
Part No.
Material
130
2 017 845
ST37
240
2 017 847
ST37
130
2 017 846
1.4571
240
2 017 848
1.4571
500
2 017 849
ST37
500
2 017 850
1.4571
Light source
Adjustment tube