23
8008250/1F7Y/V4-0/2022-06 | SICK
O P E R A T I N G I N S T R U C T I O N S | GM901
Subject to change without notice
MOUNTING
5
5.4
Fitting the flange with pipe
Important information on installing the flanges with tube
●
The axes of the flanges with tube must be aligned carefully to each other during
assembly. The angle deviation must be under 1°. Plan suitable reinforcements or
support constructions on thin-walled steel ducts.
●
On easily accessible measuring distances up to 2 m, the flanges with tube can be
aligned using a suitable auxiliary tube (for standard flange diameter 70 mm).
Fig. 2: Auxiliary tube to align the flanges on measuring distances under 2 m
Use an optical adjustment device on longer or not easily accessible measuring distances.
Fig. 3: Fitting suggestion
(1.2”)
30
(1.4”)
35
(
3”)
76
(18.5”)
470
470
(18.5”)
(2”)
50
Marking for fitting
position
Flange 60 x 8 x 6
DIN174
Steel tube 50 x 5
DIN 2391
M8