About this document
These Operating Instructions contain essential information on the function, installation,
start-up and maintenance of the FLOWSIC60 ultrasonic air velocity meter.
1.2
Main Hazards
Hazards due to explosive or combustible gases
The FLOWSIC60 measuring system may be used in potentially explosive atmospheres
according to the respective specifications.
1.3
Intended use
The FLOWSIC60 measuring system is designed for contactless measurement of air flow
velocity in explosive atmospheres.
1.3.1
Product identification
1.3.2
Operation in potentially explosive atmospheres
WARNING:
Hazards due to explosive or combustible gases
In potentially explosive atmospheres, only use the FLOWSIC60 specified for
such use.
Observe the information contained in these Operating Instructions during
installation work on running equipment.
Product name:
FLOWSIC60
Manufacturer:
SICK Engineering GmbH
Bergener Ring 27
01458 Ottendorf-Okrilla
Germany
WARNING:
Explosion danger in hazardous areas!
The front cover of the device electronics is sealed. Do not open the front cover!
Otherwise the explosion protection of the device expires and serious damage
and injury can occur.
The FLOWSIC60 is approved for use in Ex Zone 0. The device designation for
use in hazardous areas is:
Electronics:
– Ex ia [ia] I Ma
– Ex ia [ia] IIC/IIB/IIA T4 Ga
Transducers:
– Ex ia I Ma
– Ex ia IIC/IIB/IIA T6/T4 Ga
Compliance with the above requirements is the sole responsibility of the user.