Flow Rate Correction Method
This product can display a 20
°
C 1 atmospheric pressure corrected flow rate, a 25
°
C 1
atmospheric pressure corrected flow rate, a 0
°
C 1 atmospheric pressure corrected flow rate,
or the actual flow rate. In addition, the blower can be controlled so as to ensure that the flow
rate display matches the configured flow rate. The flow rate is calculated from the following
flow rate formula, and correction is performed in accordance with the selected flow rate
correction method.
Qc: Flow rate calculated under standard conditions (L/min)
Qa: Actual flow rate (L/min)
Pa: Atmospheric pressure (hPa)
Ps: Standard atmospheric pressure (1013.25 hPa)
Ta: Atmospheric temperature (
°
C)
Ts: Standard temperature (
°
C)
The cumulative flow rate is calculated by summing the instantaneous flow rate units
corresponding to the selected flow rate correction method, in 1 second units.
Communications
Sampling data can be retrieved by connecting a PC using the USB port provided on the
product. No particular communications software is prepared, but products such as
HyperTerminal can be used.
A special USB driver is required for communications. The USB driver can be downloaded from
Sibata website:
(http://www.sibata.co.jp/technology/technology-27640/)
In addition, please refer to the online manual available at Sibata website for details on
communications commands.