- 10 -
z
If the hopper is not securely fastened, it may drop during conveyance, resulting in
malfunction.
z
Never convey the sampler with the shuttle tube attached.
The glass inside may be
damaged, resulting in malfunction and injuries.
z
Do not convey the sampler with a particle size selector or an Andersen sampler
attached.
z
Do not insert anything into the flow meter suction port or exhaust port. Doing so could
cause a malfunction.
z
Do not damage the snap fastener or detachable hinge. Doing so could prevent
attaching the canopy to the main unit.
Installation
Conveyance
Always move the canopy, main unit, and folding stand separately.
z
Moving the product in the assembled state could cause damage.
Confirm that four quick clamp knobs and two mini-knobs are securely fastened in the canopy.
Always check the following to ensure the hopper is fastened to the main unit.
Ensure that the hopper is fitted into the hopper top stage fixing block and that the hole is
sealed with the sliding cover. Then tighten the hopper fixing screw.
Be sure to install the protective cap into the flow meter suction port.
The main unit may be transported tilted over in the horizontal position, but be careful not to
damage the various protruding parts. Either orient the handle side with few protrusions
downward, protect the protrusions, or wrap the unit with a blanket or other cushioning
material, or take other precautions.
Move the folding stand in the folded state.
z
Prevent the wires from getting pinched in the folding mechanism. If a wire is severed, the
main unit could fall during assembly.
Fit the hopper into
the hopper top
stage fixing block.
Seal the hole with
the sliding cover.