Memory
207
Measurement data
Saving and exporting
Concentration
Absorption / % transmission
Multi-wavelengths
Measurement datasets of these modes are saved with
<STORE>
or
AutoStore
in the measured value memory of
the spectrophotometer first (1000 locations).
The measured value memory is available via the
Measurement data memory
menu.
Here it is possible to view or filter stored measurement
datasets or export them into a PC readable file (*.csv)
(
<STORE>
).
Csv files of these modes cannot be imported back into the
photometer.
Spectrum
Kinetics
Measurement datasets of these modes are saved and
exported as a PC readable file (*.csv) with
<STORE>
.
Csv-files of these modes can be re-imported and viewed
on the photometer.
As the location for the PC readable files (*.csv) to be stored, it is possible to select for each export either the
spectrophotometer (
Internal DataB folder
) or an external memory (
USB memory
).
Later it is possible to transmit the files stored in the photometer (
Internal DataB folder
)
to a connected PC or an external memory (
USB memory
).
It is possible to transmit to a USB memory either individual files with measuring data
or all files stored in the
Internal DataB folder
.
12.2 Memorized results
12.2.1 Composition
A complete measurement dataset consists of:
- Consecutive number (is automatically assigned by the spectrophotometer).
- Date/time.
- Identification (e.g. ID or "AutoStore").
- User name.
- Measured parameter, e.g. method number, dilution, wavelength (depending on the measuring mode).
- Measured value with unit and if necessary citation form.
12.2.2 Operation on memorized results
Measurement datasets can be
•
Stored (see paragraph 12.3.3).
•
Displayed and printed (see paragraph 12.3.1).
•
Filtered, i.e. selected or hidden based on certain criteria (see paragraph 12.3.4).
•
Deleted (see paragraph 12.3.5).
•
Copied (see paragraph 12.3.6).