194
Kinetics mode
11.2 Method programming
- Press on F1 key
[New method].
It displays:
Remark:
In kinetics mode, methods are recorded under number from 4001 to 4020.
* Number
4001…4020
* Name
Name of the method (18 characters).
* Mode
Absorbance
or
Transmission
* Wavelength
320 – 1100 nm for UviLine 8100 and 9100
190 – 1100 nm for UviLine 9400
* Duration (Delay + Measure)
Total duration of the kinetics including the waiting time (Delay).
Format hh:mm:ss (hours:minutes:seconds)
* Interval
It is the time interval between two successive single measurements.
Format hh:mm:ss (hours:minutes:seconds)
<HOME>
Kinetics
- New method