65
17.5.6 Setting intervals
Command Main Body
first
character
second
character
ASCII code
Description
Value Response
I A
49H
41H
Interval time
setup
Interval time
A00: Successful
completion
E01: Command error
E02: Interval time error
Caution
Insert a comma between the entries for hours, minutes, and seconds; e.g., IA, hh, mm, ss.
17.5.7 Span adjustment/test command
Command Main Body
first
character
second
character
ASCII code
Description Value
Response
C 0
43H
30H
Disables Cal key and
command inputs.
*1*2
C 1
43H
31H
Span adjustment
with built-in weights
*3
C 2
43H
32H
Span test with built-in
weights
*3
C 3
43H
33H
Span adjustment
with external weight
C 4
43H
34H
Span test with
external weight
None
A00: Successful
completion
E01: Command
error
E02: Operation is
disabled.
E03: Cancelled by
operation
E04: Abnormal
completion
This command takes time because the balance sends back a response after an appropriate
operation is completed.
If the Cal key is disabled in Span Adjustment/Span Test of Function 1 [8. CA.], a span
adjustment and test command does not work.
*1 If the Cal key is disabled, however, the setting is reset to the function setting when the
balance is turned on again or [8. CA.] is changed directly.
*2 Span adjustment/test command will also be invalid.
*3 “C1” and “C2” commands are available only with the HTR series.