13.5 Judge by Deviation Values
13.5.1 Set two limit values by putting actual samples on the balance
−
Judge by Deviation
Values
−
1. Start limit value setting.
Press and hold the Set key for a few seconds.
When [r.SEt]is displayed, release the key.
The current reference value blinks.
2. Enter a reference weight.
Put a sample that provides a reference weight on
the balance, and press the Function key.
When a reference weight is saved, the value is
displayed. The next item will then be displayed.
3. Enter a lower limit value.
The display first indicates [L. SEt], and then the
current lower limit blinks. Put a sample that
provides a lower limit on the balance, and press the
Function key.
The error between the lower limit value and the
reference weight is displayed. Then, you can set
an upper limit.
4. Enter an upper limit value.
The display first indicates [H. SEt], and then the
current upper limit blinks. Put a sample that
provides an upper limit on the balance, and press
the Function key.
The error between the upper limit value and the
reference weight is displayed in the same way as
the setting of a lower limit. Then the balance
returns to measurement mode.
(Continued on next page.)