This function allows examination of results for long periods of time. It records
data and graph the max value on a particular date. This function helps in
quantifying the consistency of the end products.
Useful for auditing quality results over a period of time.
Note: If more than one test is performed in a day, it will only record the last
one and used that as the basis for the graph value.
Break Mode
Break mode is defined in the FGS-200 program as a percent drop from the
highest recorded max value on the test.
The percent drop can be adjusted based on the application and requirements of
the end user.
It is however advised to keep this value higher and with considerations on the
speed settings. Even though the test stand can detect the percent drop, if the
speed is higher the program may overshoot.
Values lower than 1% are prone to false activation due to possible vibration o
noise on the set up.
Break is designed for one cycle test, an optional continuance of the test if no
break is detected is possible by turning off the return position option.