The STOP Button is pressed
•
The set forece limit has been breached.
2. JOG (Jog Mode) Adjustment mode the test stand bracket will move based
on the activation of the UP and DOWN button. Unlike Manual Mode the
test stand bracket will stop as soon as the direction buttons are depressed
(momentary movement)
3. SING (Single Mode) Single operation for both direction. The test stand will
not only perform compression , but tension as well. The value for P1 and
P2 are based from which direction buttons is first pressed.
Available Set UP Options:
•
Force Limits for both direction
•
Dwell or hold time after the required distance or force had been
detected
•
Tare function.
Operation: The test stand will perform one complete set of compression
and tenstion test.
Note: The test stand speed is manually adjsuted from the speed knobs
The test stand will move to the first selected direction then afterwards
move to the opposite direction.
4. CONT (Continuous Mode) this is similar to Single mode with multiple
cycles.
Available Options
•
Force Limits for both direction
•
Dwell or hold time after the required distance or force had been
detected
•
Tare function.
•
Speed can be entered using the key pads
Operation: The test stand will perform repeated sets of compression and
tenstion test. Ideal for fatigue testing.
5. PROG ( Program Mode) There are 3 program modes set to the test stand
a. TOP – Top Load Test
b. STD – Standard Test
c. BREAK – Break Test
TOP Load test is designed for bottle testing where a required amount of force
is needed within a particular displacement.
Under this mode there are five optional test points, which can be programmed
for down or up direction.