9
M517-E077A
2 System startup and shutdown
1
Close active study
1
Click
on the Standard control
panel.
A confirmation dialog appears.
2
Click [OK].
The study is closed.
2
Close the application
program
1
Click
on the Standard control
panel.
A confirmation dialog appears.
2
Click [Yes].
The system will be automatically
s h u t d o w n a n d t h e c o n t r o l
cabinet will be turned OFF.
X-ray high voltage generator console
3
Turn OFF the X-ray high
voltage generator
Press
OFF.
ON lamp is turned off, and then
OFF lamp is turned ON.
After pressing the OFF button on the X-ray high voltage generator
console, do not press the ON button for at least 10 seconds. If you
do, the system may not operate properly.
Caution
Digital image processor
1
2
3
+5
0
-5
Shutdown