12
M517-E077A
2 System startup and shutdown
FPD Calibration from Right-click menu
Perform while the system is running.
Perform FPD calibration after the study is fi nished because the operation cannot be
performed during study.
1
Right-click on the main
image area
The right-click menu appears.
2
Select [FPD Calibration]
T h e F P D C a l i b r a t i o n d i a l o g
appears.
The following procedures are same
as auto FPD calibration in system
startup steps.