4 Fluoroscopy/Radiography
19
M517-E077A
Digital tomography (Option)
1
Verify
●
Patient information
●
Types of study
●
must be selected.
2
Select
Select APR for tomography.
3
Set up parameters
1
To select a parameter,
press [F2: Select].
2
To change the value,
press [F3: Value] or [F4: Value].
The exposure time is automatically
determined when the diagnostic table
moves to the starting point of tomography.
4
Move diagnostic table
to starting point
Press [F5: SET] until the key is highlighted.
Press [F1: Cancel] for canceling the exposure operation.
5
Exposure
Press and hold down the hand switch.
The tomography image appears on the main image area after each exposure.
Hold down the hand switch until the table stops.
When the hand switch is released during the process, the images for which the
exposures have been completed are stored on the hard disk.
Returning angle of X-ray tube unit
Press the oblique projection center switch
.
TEST
RUN
●
Layer height
●
Pitch
●
Number of exposure
(serial mode)
F1
F2
F3
F5
F4