4 Fluoroscopy/Radiography
17
M517-E077A
SPOT radiography/SERIAL radiography
Set the radiography condition prior to radiography as necessary.
Radiography
1
Verify
●
Patient information
●
Types of study
(APR cell title and radiography item)
2
Select
1
Select APR for SPOT or SERIAL
radiography.
SPOT radiography
2
Select Subdivisional Acquisition
format from [F1] - [F4].
The selected Subdivisional
Acquisition format is displayed
inverted.
3
Fluoroscopy
1
Verify that the fluoroscopy selection
switch
illuminates.
The X-ray is not exposed when the
fluoroscopy selection switch
is
not illuminated even if the foot switch
is pressed. Press the fluoroscopy
selection switch
to light.
2
Press down the foot switch.
The exposure indicator
illuminates.
The fluoroscopy image appears on the
main image area of digital image processor.
4
Expose
Press exposure switch.
SERIAL radiography
X-ray exposure terminates after images are acquired within preset time.
Acquired images are serially displayed (Auto Replay: ON) or the last
image appears (Auto Replay: OFF) after exposure.
Releasing the exposure switch stops the image acquisition during exposure.
F1
F2
F3
F4
Confirm the [IBS] and [Pulse
Fluoroscopy] are selected
on touch panel of X-ray high
voltage generator console
([IBS] lights up when ON).
IBS
Pulse Fluoroscopy
Note