2 System startup and shutdown
This operation maintains the quality of fl uoroscopy/radiography images. Be sure to perform
the FPD calibration once a day after system startup. Verify that no substance is located on
the diagnostic table during FPD calibration. Approx. 15 minutes is required to complete the
FPD calibration.
Auto FPD calibration on system startup
Perform when the system starts up.
1
Start up the system
After system startup, an FPD
calibration dialog appears.
A
on diagnostic table console
blinks, and the message appears.
2
Press and hold the [SET]
on the diagnostic table
console
The diagnostic table moves to
calibration position.
It sounds beep when the diagnostic
table arrives in calibration position.
An FPD Calibration start dialog
appears.
FPD calibration
Press [F5 : SET] to move Table to
calibration starting position.
Press [F5 : SET] to move Table to
calibration starting position.
[SET]