18
M517-E077A
4 Fluoroscopy/Radiography
DSA radiography (Option)
1
Verify
●
Patient information
●
Type of study
(APR cell title and radiography item)
2
Select
Select APR for DSA radiography.
3
Fluoroscopy
1
Confirm that the fluoroscopy selection switch
illuminates.
The X-ray is not exposed when the fluoroscopy selection switch
is
not illuminated even if the foot switch is pressed. Press the fluoroscopy
selection switch
to light.
2
Press down the foot switch.
The exposure indicator
illuminates.
The fluoroscopy image appears on the main image area of digital image
processor.
4
Prepare injector
5
Expose
Press and hold down the hand switch.
Mask images followed by live images are acquired (automatically).
The DSA image appears on the main image area of digital image processor.
X-ray exposure terminates after images are acquired within preset time.
Releasing the exposure switch stops the image acquisition during
exposure.