5-30
5.10 Waste discharge
After the maintenance of the instrument, carry out waste discharge to stabilize
emission.
1 Select
Prepare
-
Waste Discharge
on the menu bar.
Screen 5.3 Waste discharge
2 Set the sample on the spark stand and hold it by the sample holder.
3 Start the discharge by pressing
1:Start
or the start button on the spark stand.
4 After repeating the discharge specified times,
Content Analysis
screen is
displayed automatically.
Note
Below are the recommended conditions for waste discharge.
No. of emissions:
5 times
(The number can be set between 5 and 20 according to the
discharge conditions. Increase the number for poor discharge
conditions.)
Purge time:
3 sec.
Discharge condition:
SEQ1
Normal Spark
SEQ2
Combined Spark
Discharge time:
SEQ1
1,500 pulses
SEQ2
1,000 pulses
Cleaning: Cleaning
100 pulses