4 Operation
4-11
4.7
Recalibration
To calculate the recalibration coefficient, carry out the recalibration analysis.
1 Select
Analysis
-
Recalibration
on the menu bar.
2 Input the name of the analytical group in the
Group
textbox or select it by
pressing
S4:Group
key.
3 Select the name of the recalibration sample to be analyzed in the
Sample
combo
box.
Screen 4.3 Recalibration
4 Set the sample on the spark stand and hold it by the sample holder.
5 Start the discharge by pressing
1:Start
key or the start button on the spark stand.
6 Take out the sample, and change the position of the sample to be analyzed.
Then, repeat procedures 3 and 4 specified times.
7 Repeat procedures 3 to 6 until the analyses of all the recalibration samples are
completed.
8 Press
3:Cal.
Key.
The recalibration coefficients (
α, β and k values) are calculated and they are
automatically stored in the recalibration information.
In the case of one point recalibration, coefficients
α and β are not initialized. In
the case of two-point recalibration, k is initialized to 1.0.
Summary of Contents for PDA-7000
Page 2: ...This page is intentionally left blank...
Page 25: ...Chapter 1 Configuration and Function of the Instrument...
Page 36: ...1 Configuration and Function of the Instrument 1 12 This page is intentionally left blank...
Page 37: ...Chapter 2 Specification...
Page 43: ...Chapter 3 Starting and Stopping the Instrument...
Page 51: ...Chapter 4 Operation...
Page 65: ...Chapter 5 Maintenance...
Page 100: ...5 Maintenance 5 36 This page is intentionally left blank...
Page 101: ...Chapter 6 Installation...
Page 105: ...Chapter 7 Troubleshooting...