5-17
28 In order to even out the tip of the electrode that has been sharpened, recondition the tip
using discharge by following the procedure below. With this setting, discharge lasts for
approximately two minutes.
(i) Place a sample on the spark stand, select
Preparation
-
Waste Discharge
in the
menu to open the
Waste Discharge
screen, and set the condition below on
Condition 2
tab.
No. of emission
1
time
Purge time
3
sec.
Discharge condition
SEQ1
Cleaning
SEQ2
Cleaning
SEQ3
Cleaning
cleaning
Cleaning
Discharge time
SEQ1
9999
SEQ2
9999
SEQ3
9999
cleaning
9999
(ii) Start the discharge by pressing
1:Start
and wait until it is finished.
29 After reconditioning of the tip of the electrode, the total length of the electrode becomes
approximately 0.6mm shorter.
Adjust the position of the electrode again following steps 1, 2, 20 to 27.
Note
Be sure to readjust the analysis gap after reconditioning the tip of the
electrode.
Summary of Contents for PDA-7000
Page 2: ...This page is intentionally left blank...
Page 25: ...Chapter 1 Configuration and Function of the Instrument...
Page 36: ...1 Configuration and Function of the Instrument 1 12 This page is intentionally left blank...
Page 37: ...Chapter 2 Specification...
Page 43: ...Chapter 3 Starting and Stopping the Instrument...
Page 51: ...Chapter 4 Operation...
Page 65: ...Chapter 5 Maintenance...
Page 100: ...5 Maintenance 5 36 This page is intentionally left blank...
Page 101: ...Chapter 6 Installation...
Page 105: ...Chapter 7 Troubleshooting...