7
.
Serial Communications Connection Procedure
45
3
Check the send and receive
messages in the trace data file
as shown in the figure on the
right.
(In the example on the right,
Balance's command ”D07”
0Dh(CR) (single output with
stability information) is sent and
the response "S
␣
126.5947g
␣
"
0Dh(CR) (13 bytes) is received.)
Check that the response
matches the value measured by
Balance in step 4 of
7.4.2.
Executing the Communications
Sequence
.
*The above mentioned "
␣
"
indicates a space
*The contents of the response
data and parameters are
shown on the right.
For more information, refer to
10 CONNECTION AND
COMMUNICATION WITH
PERIPHERAL DEVICES
of the
Analytical Balance Instruction
Manual
(321-78327).
*The receive message varies
depending on Balance you use.
VRGT response data
Stability
information
"S" (stable)
(1 byte)
Sign
"
␣
" (+)
(1 byte)
Absolute value
"
␣
126.5947"
(8 bytes)
Unit
"g
␣
"
(2 bytes)
VRGT response parameters
Item
Size
Stability information
1 byte
Sign
1 byte
Absolute value
8 bytes
Unit
2 bytes
Summary of Contents for OMRON CJ Series