3. Hold the handset (phone) in the proper position.
Figure 76
4. Click the “Black defect correction” button.
The following message is displayed.
Figure 77
5. Click the “Yes” button. Communication starts.
Figure 78
6. The result is displayed within a minute as follows.
Figure 79
When an error occurs, the corresponding message appears. Solve the problem according to “Solution”.
•
Error 1
Error message: “There are too many black cracks.”
Cause: The light is too weak or strong.
Solution: At the proper distance from the light source (8,000 - 10,000 lux), perform black defect correction again.
•
Error 2
Error message: “There are many black cracks.”
Cause: The light is too weak or strong.
Solution: At the proper distance from the light source (8,000 - 10,000 lux), perform black defect correction again.
•
Error 3
Error message: “There are too many black cracks cannot be rectified.”
Cause: There is a hardware error.
Solution: Replace the camera unit.