- 16 -
SI 560A DIN SIZE WEIGHING INDICATOR
5-2. Simulation Calibration Mode(Calibrate without Test weight)
With this “Simulation Calibration Mode” you can make simple calibration without any
“TEST weight” This calibration mode uses “Load cells’ max capacity” and “Max Output
Rate(mV)”, so the weight adjustment degree might be less than “Test weight Calibration”.
The guaranteed resolution of this “Simulation Calibration” is 1/3,000.
5-2-1. Simulation Calibration Mode Start
Remove “CAL-BOLT” on the Front panel,
and press “CAL - LOCK S/W” inside.
When “CALIBR” is displayed, press
key and select
“SCAL”.
Press
key to enter to simulation
calibration mode.