SI 480E DIN SIZE WEIGHING CONTROLER
7-3. Digital Weighing Indicator
Display
Cause
Treatment
“
”
or
“
”
1. Load cell Error
2. Load cell cable Error
3.Load cell connection Error
4. A/D Board Error
5. If Analogue value is over 1,040,000.
※ When weigh “-“ value,
If it is over set max capa, “OVER” is displayed.
Ex) Even though set max capa is “100” and it is
over “-100”, “OVER” is displayed.
1. Under “TEST” mode 1, check analogue
value. If you cannot get any analogue value
or there is no change although adding
load, please check load cell, load cell cable,
connection conditions first.
2. Replace another load cell, and check the
indicator condition. If you have same
problem, please replace new indicator and
check A/D board error.
3. Try to connect the indicator’s A/D with the
other indicator.
4. Check the power and connection of
terminal.
“
”
1. Power is ON, when some materials are on
weighing part.
※ Under “Normal Mode”, if there are more than
20% loading of Max capacity, “Un-Pass” display
will be appeared and indicator will stay until
removing the load.
※Setting Back-up mode it can memory empty
value, and it becomes set value without
displaying ”Un-pass”
1. If you set “Normal Mode”, please check
weighing part empty or not before turn on
the power. If there are some materials
in/on weighing part, please remove those
materials and turn on the power.
2. Please try to set F102-01(Back-up) mode so
that the indicator can remember first
empty value.
“
”
When Power is on, “SET” displays.
It means EEPROM has some problem.
Please contact the distributor or Head Office.
“
”
H/W has some problem.
“
”
The dead Battery
※
“
”
on the screen relay output, Analog output, ZERO and PRINT key not activate