49
SI 480E DIN SIZE WEIGHING CONTROLER
7-2. Calibration Process
Display
Cause
Treatment
When Max capacity/digit value is over 20,000
Re-input the Max Capacity, less than 20.00
(Max Capacity / Digit)
Standard weight value is over than Max Capacity
Re-input Standard weight value with Number keys,
under Max Capacity
Standard weight value is less than 10% of
Max Capacity
Re-input Standard weight value with Number keys,
more than 10% of Max Capacity
1. Amp. Gain is too big
2. Sig+ and Sig- wire connection error
3. Test weight is not loaded
Check standard weight’s weight with set value.
If there is difference between set value and real
weight, please re-input the value
(set value is too small)
1. Amp. Gain is too small
2. Sig+ and Sig- wire connection error
3. Test weight is not loaded
Check standard weight’s weight with set value.
If there is difference between set value and real
weight, please re-input the value
(set value is too big)
Under “F-function” model, set value is “N.A”
Check the correct value and re-input
When there is continuous vibration on the weighing
part,, indicator cannot process calibration any more.
- Find vibration cause and remove
- Load cell check
- Load cell cable and connecting condition check