background image

- 4 - 

 

2. INTRODUCTION 

 

2-1. Introduction 

Thank you for your choice of this SI480E Industrial Digital Weighing Indicator. 

This SI480E model is high-performance weighing Indicator. 

Please review and learn this instruction Manual and enjoy your process efficiency   

with “SI 480E” Weighing Indicator. 

2-2. Cautions 

1. Don’t drop on the ground and avoid serious external damage on item. 

2. Don’t install under sunshine or heavy vibrated condition. 

3. Don’t install place where high voltage or heavy electric noise condition. 

4. When you connect with other devices, please turn off the power of item. 

5. Avoid from water damage. 

6. For the improvement of function or performance, we can change item specification without 

previous notice or permission. 

7. Item’s performance will be up-dated continuously base on previous version’s performance. 

8. If the equipment is used in a manner not specified by the manufacturer, the protection 

provided by the equipment may be impaired. 

9. If the operation environment is dusty, regular cleaning might be needed. 

 

2-3. Features 

1. SI 480E model is the standard 1/8 DIN SIZE and compact enough, so it is easy to install. 

2. It has wide range of DC Input. 

3. Front panel is covered with Polycarbonate film, strong against dust and water. 

4. RS-422/485 serial port standard installed. 

Summary of Contents for SI 480E

Page 1: ...Digital Weighing Indicator SI 480E Instruction Manual 2014 06 25...

Page 2: ...Load Cell Installation 10 Page 5 Set up 11 Page 5 1 Set Up mode 11 Page 5 2 TEST Weight Calibration Mode 14 Page 5 3 Simulating Calibration Mode 18 Page 5 4 F FUNCTION Setting 5 5 Test Mode 22 Page 34...

Page 3: ...y Rights 1 All Right and Authority for this Manual is belonged to SEWHA CNM CO LTD 2 Any kinds of copy or distribution without permission of SEWHA CNM CO LTD will be prohibited 3 This manual may be ch...

Page 4: ...ices please turn off the power of item 5 Avoid from water damage 6 For the improvement of function or performance we can change item specification without previous notice or permission 7 Item s perfor...

Page 5: ...Full Scale Analogue Sampling sec 60times sec MAX Environment Operating Temperature Range 10 40 14 104 Operation Humidity Range 40 85 RH Non condensing Function Calibration Mode Test Weight Calibration...

Page 6: ...unication data Print data Lamp is ON RxD When the Indicator receives Serial communication data Lamp is ON 3 2 3 Key Operation 1 Normal Mode Make Weight value as Zero 2 Calibration Mode Cancel the valu...

Page 7: ...hin 2secs enter to SET UP mode 2 Hidden Mode on pressing this key by 4 second enter to Hidden mode 2 Calibration Mode Enter 3 F Function setting Save the value go to next step Setup Mode It is a mode...

Page 8: ...3 Serial Interface terminal Stand serial port is RS485 Communication Method TX Terminal TX Terminal RX Terminal RX Terminal RS 422 Standard TX TX RX RX RS 485 Standard Not used Not used RTX RTX RS 23...

Page 9: ...ATION 4 1 External Dimension Cutting Size External Dimension unit mm Cutting Size unit mm 4 2 Installation Components SI480E Connector 3EA 3P 5P 7P Pin terminal 15EA User Manual 91 118 48 96 44 116 2...

Page 10: ...Do not weld near the load cells Indicators or other devices Load Cell Installation 1 You can connect Max 8pcs of same capacity Load cells at once 350 2 You have to make horizontal balance on the grou...

Page 11: ...SET UP is displayed SETUP Mode is activated 5 1 2 Start SET UP Mode Password Use Refer Hidden Option HF07 Press key four times within 2sec If P W displays input 4 characters password If Password is r...

Page 12: ...imulation Calibration key during 4 seconds Pass word F FUNCTION Mode key 4 times Test Mode1 Analog Value key 4 times Display key 4 times Key test key 4 times Test Mode2 Serial I F key 4 times Standard...

Page 13: ...replace LOAD CELL or Indicator you have to Calibration process once again When you start calibration mode TARE HOLD PRINT will be reset Before processing calibration please warm up the indicator duri...

Page 14: ...screen input PASSWORD press key On the screen SET CAL press key On the screen CALIBR press key Go into CALIBRATION MODE 5 2 2 Setting Capacity of weighing Scale When CAPA is showed input max capacity...

Page 15: ...with key finally press key to save the data Max Decimal point will be 0 001 and digit can be selectable among 1 2 5 10 20 50 Digit and Decimal point must be fulfill the below condition Division value...

Page 16: ...ation CAL00 CAL09 will be operated twice when resolution Division value Max capacity value is less than 1 10 000 At this step if there is some force or vibration on weighing scale and unstable conditi...

Page 17: ...ond automatically To guarantee the preciseness SPAN calculation CAL00 CAL09 will be operated twice when resolution Division value Max capacity value is less than 1 10 000 After calculation span value...

Page 18: ...cells max capacity and Max Output Rate mV so the weight adjustment degree might be less than Test weight Calibration The guaranteed resolution of this Simulation Calibration is 1 3 000 5 3 1 Simulatio...

Page 19: ...ach load cell s capacity and make setting with max capacity Ex If there are 4pcs of load cells and each load cell s Max capacity is 1 000kg Then total Max Capacity will be 4 000kg 1 000 x 4 and you ha...

Page 20: ...o guarantee the preciseness DEAD weight calculation CAL00 CAL09 will be operated twice when resolution Division value Max capacity value is less than 1 10 000 5 3 5 Input Max Output Rated Output Volta...

Page 21: ...ration process In case of multiple pieces of load cells are connected the rated output will be same as single load cell Because plural load cells are connected with parallel connection the sum of rate...

Page 22: ...optimal operation of SI 480E controller Starting F FUNCTION Mode Press kye 4 times Displaying SETUP press key 101 00 ESC LEFT RIGHT INCREASE SAVE Set the function no set value and Press key Whenever p...

Page 23: ...4 Display Up Date Speed 09 01 Slow 1time per 1sec 09 Fast 60times per 1sec 108 Buzzer sound External input detection 00 00 Buzzer sound 01 No Buzzer sound 110 Weight Unit 00 00 kg 01 g 02 ton 111 Lang...

Page 24: ...te 00 00 Disuse 01 Use 212 Tare Delay Time 00 00 Disuse 01 10 Use Unit 1 sec 213 Auto tare set when weighing starts 00 00 Disuse 01 Use 214 Tare Removal Timing 00 00 Manual 01 Auto at empty range 02 A...

Page 25: ...emoval 08 Print 09 Subtotal Print 251 Zero state lamp output standard 00 00 Near Zero 01 Zero 301 Parity Stop bit 00 00 Data bit 8 Stop bit 1 Parity bit None 01 Data bit 8 Stop bit 1 Parity bit Odd 02...

Page 26: ...on every steady state 02 Single time finish weighing process 03 When input PRINT key 307 Modbus Transmit Data MSB LSB location 00 00 Standard 01 Change 352 Print Format Setting 00 00 Continuous Print...

Page 27: ...nish 04 Manual Auto At every steady states 05 Manual Auto At the first steady states Display Up Date Speed 104 09 01 09 01 Slow 1 time per 1 sec 09 Fast 60 times per 1 sec Buzzer sound External input...

Page 28: ...ays active 01 Active under steady condition only Zero key Operation Range 209 00 Active within 2 of Max Capacity 01 Active within 5 of Max Capacity 02 Active within 10 of Max Capacity 03 Active within...

Page 29: ...ld mode 216 00 Sample Hold Hold current weight until Hold Reset 01 Peak Hold Measure Max weight value and hold on display 02 Average Hold Hold average value Hold delay time 217 00 00 10 00 Disuse 01 1...

Page 30: ...Hold Hold removal 02 Tare 07 Hold Hold removal 03 Tare removal 08 Print 04 Tare Tare removal 09 Subtotal print Zero state lamp output standard 251 00 Near Zero 01 Zero Parity Stop bit 301 00 DATA Bit...

Page 31: ...03 MODBUS RTU Command mode Check Sum detection selection F303 01 304 00 Disuse 01 Use Stream mode DATA Transference Format selection Refer chapter 6 1 4 305 00 Format 1 19byte 01 Format 2 22byte 02 F...

Page 32: ...thod F function 103 Print input Key Comm External input Printing out data Saving Data 00 Manual Current weight Current weight 01 Auto At every steady states Recent Stable weight Steady weight Steady w...

Page 33: ...key F LIST Subject Default Contents HF01 Serial Number Check HF03 S W Version Check Program version check HF04 H W Version Check HF05 DATE Y M D Check Able to modify HF06 TIME H M S Check Able to mod...

Page 34: ...ess key Cancel Go back Press key If there is no change although pressing keys or loading some force on in weighing part it may something wrong with load cell cable connector or A D board TEST MODE 1 E...

Page 35: ...og value to 0 Display from 1 million unit Display from ten million unit 5 5 2 Display check mode Test FND and LED Blink FND and LED by 1 Segment gradually 5 5 3 Key check mode Press each key on the sc...

Page 36: ...e and check the transference and receipt At the normal operation display will be blinked TESTING PROTOCOL Format STX Id No TEST ETX You cannot test Standard and Extended Serial Interface at the same t...

Page 37: ...ce RS 485 Standard installed Selectable 6 1 2 Serial Interface RS 422 Standard Selectable 5 1 4 8 0 e 5 1 4 8 0 e 5 1 4 8 0 e 5 1 4 8 0 e 5 1 4 8 0 e 5 1 4 8 0 e 32 pcs of Indicators are connectable 3...

Page 38: ...ated place from Power cable or other electric cables and wires and please use shielded cable for better performance 6 1 4 Data Format 1 Format 1 ID Number is not be transferred Refer F305 00 19 byte H...

Page 39: ...UNSTEADY 3 Format 3 ID Number State Refer F305 02 17 byte 4 Refer F305 03 22 byte LAMP DISPLAY Bit 7 Bit 6 Bit 5 Bit 4 Bit 3 Bit 2 Bit 1 Bit 0 1 0 1 1 1 1 1 1 1 STEADY 1 Hold Print Gross Weight TARE...

Page 40: ...ID RCWT ETX 22 byte Current data STX ID RCWD ETX 46 byte Current date data STX ID RDAT ETX 14 byte Current time data STX ID RTIM ETX 14 byte Tare weight STX ID RTAR ETX 18 byte 6 1 7 Write Command Su...

Page 41: ...te1 O Over Load S Steady U Unsteady State2 N Net weight G Gross weight Ex Steady S TARE not used N 0 000kg Indicator memory data ASCII STX ID 2byte RCWD ETX HEX 02 30 31 52 43 57 44 03 Response STX ID...

Page 42: ...10 000kg Current Time data ASCII STX ID 2byte RTIM ETX HEX 02 30 31 52 54 49 4D 03 Response STX ID RTIM Current Time 6byte ETX 12 00 00 Current date data ASCII STX ID 2byte RDAT ETX HEX 02 30 31 52 4...

Page 43: ...X HEX 02 30 31 57 48 52 53 03 Response Normal STX ID ACK ETX Error STX ID NAK ETX PRINT Data will be transferred to the port which is set as print mode Function 303 310 02 ASCII STX ID 2byte WPRT ETX...

Page 44: ...um Sum the value from STX to ETX and converts to ASCII 2byte and transfer Convert the Sum value HEX to ASCII and transmit 28byte ex The sum HEX value from STX to ETX 02 30 31 52 43 57 54 03 is 1A6h Th...

Page 45: ...al Key Modbus Function Codes 03 0x03 Read Holding Registers 04 0x04 Read Input Registers 06 0x06 Write Single Registers 16 0x10 Write Multiple Registers CRC Check Method is CRC 16 Ad Length Feature De...

Page 46: ...are OUT1 OUT2 OUT3 OUT4 Hold 3 Error register Address 21 Length 2 0 1 2 Load cell Error Over Load Set point Error 4 Key register Address 441 Length 1 0 1 2 3 4 5 6 7 8 9 10 11 Start Stop Zero Tare Tar...

Page 47: ...format is already programmed and fixed with SE7200 7300 model 6 2 1 Printing Format Using the RS 485 or 422 interface please use convertor and converts to RS 232 and connect with Serial printer If you...

Page 48: ...G is about 350 3 5 3 Isolate Resistance is more than 100M Weight Value is increased regular rate but not return to Zero 1 Load cell Error 2 Load cell connection Error 1 Check Load cell connection 2 Me...

Page 49: ...or 3 Test weight is not loaded Check standard weight s weight with set value If there is difference between set value and real weight please re input the value set value is too small 1 Amp Gain is too...

Page 50: ...ect the indicator s A D with the other indicator 4 Check the power and connection of terminal 1 Power is ON when some materials are on weighing part Under Normal Mode if there are more than 20 loading...

Page 51: ...EWHACNM permission Any kinds of Mal function Defection or External damage caused by operator Any kinds of Mal function Defection caused by using spare part from Non Authorized Distributor or Agent Any...

Reviews: