SI 400 DIGITAL INDICATOR
5-1-3. Decimal point and division setting
Max decimal point will be 0.001, and digit can be selected among 1, 2, 5, 10, 20, 50. Digit
and decimal point must be fulfilled under the below condition.
- (division value / Max capacity value) cannot be over 1/20,000.
If this condition is not fulfilled, “
Err-1
” will be displayed and move back to capacity setting
mode.
5-1-4. Measuring the “DEAD” Weight of Weighing Scale.
After “DIVI” is displayed, locate the decimal point with
and
keys, and set the
division with
and
keys. Press
key to save.
When
“DEAD”
displays, Press
key, then indicator will calculate dead weight of scale part
automatically (While this process, there should be nothing on the scale part.