SI 400 DIGITAL INDICATOR
DATE(Y,M,D) Check / Modification
HF05
Check the date or adjust when it is wrong.
TIME(H,M,S) Check / Modification (24Hours)
HF06
Check the time or adjust when it is wrong.
Password Setting (4digit)
HF07
Password is required when you enter to hidden function.
Enter the password twice.
1 2 3 4 5 6 7 8 9 0
Password combination within 0~9
Maximum Capacity Weight Check and Modification
HF08
Check the max capacity which is set under calibration mode.
Division
HF09
Check the division which is set under calibration mode.
span value
HF10
Check the weight of test weight which is used for your last calibration.
Zero Analog
HF11
Check the analog value of ZERO.
Analog Output Use Setting
HF13
● 00 4-20mA Output
01 0-10V Output
Minimum Analog Output Setting
HF14
Minimum Analog Output (Analog out 4~20mA / 0~10V).
key press (–) Setting. Input range : -20 ~ +20 , basic value : 0
Maximum Analog Output Setting
HF15
Maximum Analog Output (Analog out 4~20mA / 0~10V).
key press (–) Setting. Input range : -20 ~ +20 , basic value : 0
Function List Factory Reset
HF16
Change to default F-setting
Simulation calibration Setting
HF30
00 : Discuse, 01: Use