- 13 -
SI 400 DIGITAL INDICATOR
5. SET-UP
5-1. Test Weight Calibration Mode (Using test weight)
5-1-1. Start Test Weight Callibration Mode
Remove “
CAL-BOLT
” on the Rear panel, .
and press “
CAL - LOCK S/W
” inside.
When “
CALIBR
” displays, press
key.
select
“WCAL”
and press
key.
5-1-2. Setting “Capacity of weighing Scale”
If you want to set Max capacity as 1,000kg and the division is
0.1 (100g), then just input “1000”.
key for going back to zero,
key for gradual decrease from unit digit.
After displaying “
CAPA
”, input max capacity with keys & Press
key to save & move to
next step.