15
Velocity, Flow, Pressure, ACH Selection
User Selectable for Velocity, Flow, Pressure or
Air Changes per Hour. Whichever option is
chosen in this menu will be shown as the default
in the secondary menu under the
V/Flow/Pressure section. Choose “enter” to
activate the V/Flow/Press menu. The menu will
flash and then press the “down arrow” to select
desired mode. Press “enter” to select.
Alarm Delay
Set the desired time for Alarm Delay. Choose
“enter” to activate the Alarm Delay menu. The
menu will flash, then press the “down arrow” to
select correct digit: “enter” moves to next digit.
Mute Timeout
Sets the desired amount of time to engage Mute
Timeout. Choose “enter” to activate the Mute
timeout menu. The menu will flash, then press
the “down arrow” to select correct digit: “enter
moves to next digit.
Buzzer Enable
Enables or disables the audible buzzer alarm.
Choose “enter” to activate the Buzzer Enable
menu. The menu will flash, then press the “down
arrow” to select enable or disable. Press “enter”
to select.
Door Switch
Door switch enable/disable. Enable if used with
a door switch to detect if door is open.