10
Step 5.
Press the
HOLD/mode
key to display the number "0".
Note:
The decimal point that was set during calibration (CAL_U, CAL_C) will be
displayed.
Step 6.
Set the value you wish the meter to display at the presently applied pressure.
Press the
SCAN/adj
key to move the numbers up (+).
Press the
HOLD/mode
key to move the numbers down (-).
When the desired number is reached, press the SCAN/adj and HOLD/mode keys
at the same time. The number will flash repeatedly. Then, press the SCAN/adj key
to readjust the number or the HOLD/mode key to accept it. After the number is
accepted, the meter will display "bUSY" for 10 seconds and then resume normal
operation"