SERVOTOUGH LaserExact Monitor, Operator Manual 02950001A_1
ELECTRICAL
CONNECTIONS
55
set “NC”, normally closed. This may be changed to “NO”, normally open with dip
switches marked “Rel ‘x’ NC/NO” on the AUX board.
Note that external relays connected to the instrument may induce current or
voltages that exceed the above ratings. If in doubt external protection diodes must
be implemented to protect the internal instrument relays.
The PLC/PLS (Programmable Logic Controller) 4-20 mA outputs are of ACTIVE
type and isolated electrically from the electronics unit’s electronics. Maximum load
is 500
Ω
.
There are up to 3 different 4-20 mA outputs available. Per default only the outputs
for gas concentration are implemented.
For single gas instrument the outputs are configured as follows:
PLC1
– output for gas concentration,
PLC2
– optional output for transmission (4mA=0%, 20mA=100% transmission),
PLC3
– not used.
For dual gas instrument the outputs are configured as follows:
PLC1
– output for gas-1 concentration,
PLC2
– output for gas-2 concentration,
PLC3
– optional output for transmission (4mA=0%, 20mA=100% transmission).
For instrument measuring gas and temperature the outputs are configured as
follows:
PLC1
– output for gas concentration,
PLC2
– output for spectral temperature,
PLC3
– optional output for transmission (4mA=0%, 20mA=100% transmission).
6.4
RS-232 and Ethernet connections
Junction list
Terminal:
Description:
Comments:
J13
1
Data Carrier Detect
Connected to 4.
J13
2
Transmitt Data
J13
3
Receive Data
J13
4
Data Terminal Ready
J13
5
Ground
J13
6
Data Set Ready
Connected to 4.
J13
7
Request To Send
Connected with 8.
J13
8
Clear To send
Connected with 7.
J13
9
Ring Indicator
Not Connected.
Table 6-3: Cabinet – RS 232 connection
Alternatively connector J15 may be used (remember gender changer).
Summary of Contents for SERVOTOUGH LaserExact 2950
Page 2: ...This page intentionally blank...
Page 6: ...SERVOTOUGH LaserExact Monitor Operator Manual 02950001A_1 IV This page intentionally blank...
Page 69: ...SERVOTOUGH LaserExact Monitor Operator Manual 02950001A_1 APPENDIX 63...
Page 70: ...SERVOTOUGH LaserExact Monitor Operator Manual 02950001A_1 64 APPENDIX...
Page 71: ...SERVOTOUGH LaserExact Monitor Operator Manual 02950001A_1 APPENDIX 65...