10. Workshop Files
The workshop files are intended to enable a quick start of measurements
with a (frequently used) base setup. By means of the main menu entries
File
→
Save Workshop File
and
File
→
Load Workshop File
the current state
of the software can easily be saved and restored in the
swsf
workshop file.
The following settings/states are stored and loaded:
•
Current
Sample List
•
Current
Macro List
•
Location of last known
Reference Measurement
.
Automatic impedance masks positioning:
In a laboratory environment it may happen that the setup partially changes
from time to time. For example impedance measurements with predefined
tolerance masks require a constant measurement setup with respect to the
position of the mask position - this position varies with the length of the
used cable.
A position-detection algorithm allows for an automatic adaption of the po-
sitions of the tolerance masks upon loading a workshop file.
Please note:
to be able to use this functionality correctly, it is
very important
that a current
Reference Measurement
has been made.