3.5.3 Result Visualization Window
Figure 3: Impedance measurements results
General Remark
Most
Measurement Modes
(see chapter
) yield more information than can
be displayed in the yellow result field of the
Main Window
. To display the
specific results more in detail the
Result Visualization Window
is used. The
different kinds of visualizations depend on the currently selected measure-
ment mode and are therefore described in the specific sections.
TTC/12TC Calibration
The software allows the correction of systematic errors by means of OPEN,
SHORT, LOAD calibration, referred to here as
Three Term Correction (TTC)
.
An additional THRU is used for a full 2-port calibration (
12-Term Correc-
tion = 12TC
). The
TTC/12TC
can be activated/deactivated using the
TTC ac-
tive
checkbox. Please note that deactivation is not allowed for frequency
domain measurements and therefore the button is disabled in this case.
19