Lacto Spark

Virtually

Painless

Lactate Measuring System

Lacto Spark

USER MANUAL

Summary of Contents for Lacto Spark

Page 1: ...Lacto Spark Virtually Painless Lactate Measuring System Lacto Spark USER MANUAL...

Page 2: ...onsult accompanying documents Consult instructions for use Biological risk Catalog number Temperature limitation Manufactured by About this Manual This manual is for the Lacto Spark Meter Throughout t...

Page 3: ...rip 6 3 Preparing to test your Blood Lactate 7 Lancing device handling procedure 7 Performing a blood Lactate test 9 Applying blood to Lacto Spark test strip 10 Strip Ejection 11 Ejection of used lanc...

Page 4: ...st strips with the meter can produce inaccurate results Do use the Lacto Spark system for Testing blood lactate Testing only with fresh capillary whole blood samples Warning The Lacto Spark system con...

Page 5: ...r if the system check screen does not match the example here Yes Button Previous Reading NO Button Exit Test Strip Part Insert the top end of a new Lacto Spark Test Strip here The meter powers on when...

Page 6: ...tell you when the meter is ready to apply blood Battery Symbol Strip Symbol This symbol works with Blood Drop symbol in an animation to indicate that the meter is ready to apply blood Display Displays...

Page 7: ...em has an operating range of 10o C to 40o C Store the test strip package in a cool dry place between 10o C and 30o C Use test strips only within the system operating temperature range Keep away from d...

Page 8: ...e package before starting the test Note To ensure accurate results wash your hands thoroughly with soap and water Make sure there is no cream or lotion on the test site Dry your hands and test site th...

Page 9: ...the lancet fits into the lancet holder Push the lancet into the device until it snaps into place and is fully seated in the holder Remove the protective layer cover of lancet Step 3 Replace the lancin...

Page 10: ...op of blood forms on your fingertip Apply the blood sample on the strip connected to meter Performing a blood Lactate test Insert test strip to turn On the Meter Insert and slide a Lacto Spark test St...

Page 11: ...squeeze or massage your finger tip to obtain a drop of blood Apply Blood Sample on the strip The Lacto Spark Test Strip acts like a sponge and draws the blood into the strip through the sample area W...

Page 12: ...o small or strip might be faulty Replace the new strip The test result is complete when your blood lactate test result is shown on the display screen and you hear beep sound twice On average your read...

Page 13: ...wise and then pulling it straight off of the device Cover the exposed lancet tip Before removing the lancet place the lancet protective cover on a surface then push the lancet tip into the cover Eject...

Page 14: ...iry date of the strips and control solution has not elapsed See product insert sheets for details the control solution test results should fall within the range of results printed on the label of the...

Page 15: ...s on and off repeatedly until sufficient control solution has been added to the test strip Warning Blood samples and blood products are potential sources of hepatitis and other infectious agents Handl...

Page 16: ...lling the locking catch upwards and remove the battery compartment cover as shown in the image Battery removal Remove the battery by tilting the meter to sideways Replacing the battery Place new batte...

Page 17: ...er immerse the meter in water or any other liquid The meter should be handled with care Dropping rough handling etc may damage the meter Also protect the meter from moisture prolonged direct sunlight...

Page 18: ...incompletely inserted into the meter Insert the Lacto Spark Test Strip Place the end of the strip with the two contact bars as far into the meter as it can go Test does not start after applying the b...

Page 19: ...ngs Operating Relative Humidity 5 to 90 non condensing Operating Temperature 10o C to 40o C Power Source One CR 2032 3 volt lithium battery replaceable Result Range 0 2 to 25 0 mmol L Sample Capillary...

Page 20: ...SENSA CORE MEDICAL INSTRUMENTATION PVT LTD Plot No 3 EPIP Zone Pashamylaram Sangareddy Dist Hyderabad Telangana INDIA 502307...