It is important to use a new lancet each time you obtain a blood sample. This will
help to prevent from infections. Discard the used lancet carefully after each use to
avoid unintended lancet stick injuries. Used lancets may be considered
biohazardous waste in your area. Be sure to follow your healthcare professional’s
recommendations or local regulations for proper disposal.
4. Testing Control Solutions
NOTE:
Before running a blood or control test, check that the expiry date of the
strips and control solution has not elapsed. (See product insert sheets for details.)
the control solution test results should fall within the range of results printed on
the label of the control solution. You should run a control solution test:
Before using your Lacto Spark Meter for the first time and at least once a week
thereafter.
Each time you open a new box of Lacto Meter Test Strips
If you leave the Lacto Spark Test Strip vial cap open for an extended period of
time
If you drop your Lacto Spark Meter
If your results are higher or lower than expected
User should follow the appropriate Indian, State Guide Lines concerning the
running quality controls.