SENS IQ Technical Manual
32
Service Hotline: 1.800.742.2326
1.303.678.7500
DOES CHARGER
OUTPUT MATCH
BATTERY
VOLTAGE?
CHANGE CHARGER TO MATCH
BATTERY VOLTAGE
CHECK SEMICONDUCTORS
AND OUTPUT CAPS
BREAKER STILL
TRIPS?
SEMICONDUCTORS
AND OUTPUT CAPS
OPERATING
PROPERLY?
REPLACE
SEMICONDUCTORS/OUTPUT
CAPS
BREAKER STILL
TRIPS?
START
N
N
Y
Y
N
N
Y
Y
REPLACE BREAKER
BREAKER STILL
TRIPS?
CONTACT
CUSTOMER SERVICE
Y
N
END
END
END
Figure 15: Troubleshooting flowchart for DC output breaker trip