User Manual
Z-SG2 / Z-SG2-L
ALL RIGHTS RESERVED. NO PART OF THIS PUBLICATION MAY
BE REPRODUCED WITHOUT PRIOR PERMISSION.
www.seneca.it
Doc: MI-005370-EN
Rev. 1.0.0.2
Page 46
ADDRESS
(4x)
ADDRESS
OFFSET
REGISTER
DESCRIPTION
W/R
TYPE
40045
44
ADVANCED ADC
SPEED
Configures the ADC sampling
speed.
For further information, please
refer to 10.2.9.3
It is active only if level 7
filtering (advanced) is set.
RW**
UNSIGNED
16 BIT
40046
45
ADVANCED
MOVING
AVERAGE
Represents the number of
samples in the moving
average from 1 to 100
It is active only if level 7
filtering (advanced) is set.
RW**
UNSIGNED
16 BIT
40047
46
ADVANCED ADC
CHOPPING
Enables or disables the ADC
Chopping mode.
For further information, please
refer to 10.2.9.2
It is active only if level 7
filtering (advanced) is set and
in calibration mode with
sample weight..
0 = It disables ADC Chopping
1 = It enables ADC Chopping
RW**
UNSIGNED
16 BIT
40048
47
AUTOMATIC TARE
RESET
0 = It disables the tare tracker
If >= 1 is the value of ADC
points within which to reset
the tare automatically.
If after 5 seconds of stable
weighing condition the ADC
value of the net weight
deviates by less than this
value then a new tare is
acquired.
RW**
UNSIGNED
16 BIT
40049
48
THRESHOLD
HYSTERESIS
[kg/g/t…]
It is the hysteresis value for
the threshold exceeded alarm
(MSW)
RW**
FLOATING
POINT 32
BIT