User Manual
Z-SG2 / Z-SG2-L
ALL RIGHTS RESERVED. NO PART OF THIS PUBLICATION MAY
BE REPRODUCED WITHOUT PRIOR PERMISSION.
www.seneca.it
Doc: MI-005370-EN
Rev. 1.0.0.2
Page 44
ADDRESS
(4x)
ADDRESS
OFFSET
REGISTER
DESCRIPTION
W/R
TYPE
40011
10
DIGITAL INPUT
TYPE /
CALIBRATION
MODE
(MSB) DIGITAL INPUT TYPE:
Selects the operation of the
digital input
0 = The digital input acquires
the tare
1 = The status of the digital
input is shown on Modbus
without effects
(LSB) CALIBRATION MODE:
Sets the type of operation and
calibration
0 = It uses factory calibration
1 = It uses the calibration with
sample weight
RW**
UNSIGNED
16 BIT
40029
28
CELL SENSIBILITY
[mV/V]
Ratio value of the load cell
sensitivity in [mV/V] (MSW)
RW**
FLOATING
POINT 32
BIT
40030
29
Ratio value of the load cell
sensitivity in [mV/V] (MSW)
RW**
40031
30
CELL FULL SCALE
[kg/g/t…]
Full scale value of the load
cell in technical units (MSW)
RW** FLOATING
POINT 32
BIT
40032
31
Full scale value of the load
cell in technical units (LSW)
RW**
40033
32
STANDARD
WEIGHT VALUE
[kg/g/t…]
Value of the standard weight
in technical units to be used in
the calibration mode with
sample weight (MSW)
RW**
FLOATING
POINT 32
BIT
40034
33
Value of the standard weight
in technical units to be used in
the calibration mode with
sample weight (LSW)
RW**
40035
34
OUTPUT WEIGHT
STOP SCALE
[kg/g/t…]
Full scale value of the net
weight for the analogue output
in technical units (MSW)
RW**
FLOATING
POINT 32
BIT
40036
35
Full scale value of the net
weight for the analogue output
in technical units (LSW)
RW**
40037
36
OUTPUT WEIGHT
STOP SCALE
[kg/g/t…]
Start scale value of the net
weight for the analogue output
(MSW)
RW**
FLOATING
POINT 32