MANUAL DE INSTRUCCIONES CODIGO 80196 REV B Octubre 2019 (Sujetas a modificaciones sin previo aviso) Pag.: 40
J
.P. SELECTA s.a.u.
Autovía A2 Km 585.1 Abrera 08630 (Barcelona) España
Tel 34 937 700 877 Fax 34 937 702 362
e-mail: selecta @jpselecta.es - website: http://www.jpselecta.es
9 Technical specification
Mains voltage:
230V 50/60Hz
(See plate)
Other power supply voltages could be
available (Ask to your dealer)
Electrical power:
600W (6 places) 200W ( 2 places)
Maximum comsuption:
3 A
1.5 A
Fuses:
4 A fast. dimensions: 20x5 mm
Power supply cord:
Phase, Neutral, Earth (section 1mm
2
)
Weight (Net):
25 Kg
19Kg
Dimensions: W x L x H
400x750x700 mm 400x450x700 mm
Thimbles size
External diameter 26mm /height 60mm
Max working temperature
80 to 240ºC
Solvent recovery:
Better than 60% - 80%
Reproducibility:
Relative Standard deviation)< 1%
Sound level:
Lower than 40 dBA
Protection degree:
IP55
Ambient conditions:
Inner use.
Altitude up to 2000m
Temperature from 5ºC to 40ºC.
Relative maximum damp 80% for above
31ºC temperatures, decreasing linealy to
50% for 40ºC.
Overvoltage category:
Category II
Pollution degree:
2
Glass material:
Borosilicate 3.3
Carriage material:
Iron with overlap with RILSAN coating
chemical resistant.
Hosepipe material:
Silicone.
Number of programs:
15
“Boiling” extraction time:
from 0 to 999 min
“Rinsing” extraction time:
from 0 to 999 min
Solvent recovery time:
from 0 to 99 min
10 Warranty
The warranty covers the defective components or faulty assembling of the unit for 1
year since the purchasing date.
Keep purchasing bill as the only valid document for the purchasing date.
The warranty does not cover the damages caused for the unit malfunction. The unit
must be suitably qualified before being used for valuable samples processing.
Return goods are not accepted without previous J.P. SELECTA, s.a.u. approval.
ERROR 7:
SOME PROBE IS ABOVE 250ºC
Causes: Probe damaged, triac of the shorted probe.
Action: The device warns and goes to rest immediately
ERROR 8:
IN OFF STATE AND WITH A MIDDLE TEMPERATURE LESS
THAN 70ºC, SOME PROBE IS ABOVE 80ºC
Causes: Probe damaged, triac of the shorted probe.
Action: The device warns and goes to sleep after a few seconds