MANUAL DE INSTRUCCIONES CODIGO 80196 REV B Octubre 2019 (Sujetas a modificaciones sin previo aviso) Pag.: 30
J
.P. SELECTA s.a.u.
Autovía A2 Km 585.1 Abrera 08630 (Barcelona) España
Tel 34 937 700 877 Fax 34 937 702 362
e-mail: selecta @jpselecta.es - website: http://www.jpselecta.es
To run a program:
• Open the water tap for coolant water.
• Select the menu «OPERATION» to run an stored program. See 5.2.
- Bear in mind that program number “0” is only for checking purposes.
- After «OPERATION» is selected, the first program shown on the display is the
last program used or edited.
- To start a program, press key ENTER/START and could be stopped at any time
by pressing key ESC/STOP.
• The program initiates the «BOILING» step, so place the thimbles on the low posi-
tion (See 5.5). Once the «BOILING» step finished, the unit makes some BEEP and
the display flashes.
Do not let the unit completely unattended.
Pay attention to the end of each step to change, at time,
the extraction thimbles position.
• The unit jumps directly to «RINSING» step. Move the thimble to the up position.
Once the «RINSING» time finishes, the unit makes some BEEP and the display flashes.
• The unit jumps directly to «RECOVERY» step. Close the PTFE manual valves for
solvent recovery. Once the «RECOVERY» time finishes, the unit makes some BEEP
and the display flashes.
BOILING 20m t85 1
19:40 57,4
Program step
Set time
Set temperature
Remaining
time mm:ss
Actual tempe
-
rature
Heating status:
0: Stopped
1: Active
Display when a program is running
Automatically apears the parameters
for the last program used (edited or
run).
The display flashes between them.
OPERATION
Use key ENTER
to run the extraction
program.
Press ENTER to entry into menu
OPERATION.
Use the key ESC
to return to pre
-
vious step.
SOLVENT
DIETHILIC ETHER
PROGRAM
01 XXXXXXXX
Bo Ri Re TºC
10 10 10 80
SOLVENT
ACETONE
PROGRAM
02 XXXXXXXX
Bo Ri Re TºC
10 10 10 140
Use keys (Up/Down) to select the program to run.