11 Error messages
12 Spare parts
10044 R4 closing
13006 Contactor
15410 Fuse 16 A
16057 Electrovalve NA 3 mm 24 V
16212 Pressure switch
20017 Bipolar switch
29487 Frontal circuit
29499 Power circuit
36027 Solid relay 25A
39001 Heating element 50L + 80L
43031 Thermostat
43071 Pt100 probe
21193 Gasket 50L
21271 Gasket 80L
21206 Gasket 150L
*When requesting spare parts it is mandatory to send the model and
serial number of the equipment.
10
Code
Name
Description
01
MIN TEMPERATURE Pt100 probe failure.
Stop the equipment and check the Pt100 probe connection.
02
MAX TEMPERATURE Pt100 probe failure.
Stop the equipment and check the Pt100 probe connection.
03
SECURITY RELAY
External alarm. Pressure switch or safety thermostat trip.
Stop the equipment. Allow the safety thermostat to cool down and reset
by pressing button 10 (see picture attached). The thermostat is set at the
factory, it must not be manipulated except for its resetting.
05
DOOR OPEN
Door open. Stop the equipment. Let it cool down, make sure the pressure
is 0 bar. Check if the closing system is working correctly. If the error is re
-
peated, contact the technical service.
06
UNDERTEMP
Temperature out of limits during sterilization.
The sterilization may not be correct. Check the material once the cycle is
finished. Make sure that the autoclave has enough water and that the ma
-
terial to be sterilized does not touch the temperature probe at the bottom
of the chamber.
08
OVERTEMP
Temperature out of limits during sterilization.
The sterilization may not be correct. Check the material once the cycle is
finished. Make sure that the autoclave has enough water and that the ma
-
terial to be sterilized does not touch the temperature probe at the bottom
of the chamber.
09
POWER FAIL
Network voltage drop during the cycle.
The equipment has run out of power during the operation of a cycle. Make
sure you have not turned off the equipment before the end of the cycle or
that you have had a power cut during the cycle.
11
PURGE
It can be due to different reasons: the resistance is in discomfort (check);
the purge solenoid valve is closed (check); the equipment is located more
than 1500m high. In this last case, contact the technical service.
12
DISCONNECTED
Screen and power have lost communication.
Location of the safety thermo
-
stat reset system