10 Maintenance
10.1 Temperature calibration
If a correction needs to be done, do as follows:
- Press the button (8) «ºC» and (9) «clock» simultaneously for more
than 5 seconds.
- While continuing to press these two keys, press keys (10) and (11) to
modify this parameter. For fast forward press and hold the arrow (10 or
11) for a few seconds.
Note: Correction range from -9.9 to + 9.9ºC. The modified value is added
or restored to the measured temperature by correcting it.
10.2 Safety thermostat
8. Safety thermostat regulation screw (grey).
9. Safety thermostat reset button (white).
10.2.1 Rearming the safety thermostat
Safety thermostat is an element to prevent the temperature from exceed
-
ing a value. If by a fortuitous event the thermostat is triggered, that is,
it opens the power circuit, the heating element will stop working. Usually
there is an indicator light, or an error message, that indicates that the
thermostat is in this position.
If the thermostat is triggered, check that the equipment looks good and
it is not overheating at any point.
To change its status, simply press the white button (9), and you will hear
a “click” that indicates that the thermostat has been reset. If the equip
-
ment has not cooled down sufficiently, the thermostat cannot be reset,
and you must wait for the temperature to drop before it can be reset.
10.2.2 Adjusting the safety thermostat
In some equipment such as ovens or baths and with certain processes, it
may be useful to adjust the thermostat trigger value, but for the rest of
equipments this operation should NEVER be carried out. From factory, the
thermostat is adjusted to its maximum temperature.
10.3 Gasket
It is necessary to check that the gasket is in good condition. With the
door open, remove the gasket and check that it is not deformed, has
cracks or is broken. To check it, remove the gasket from the cover, dry it
with a cloth and dry the groove of the cover where the gasket is located.
Reassemble and do a test cycle. To ensure a good seal with a use of 1
cycle a day, it is recommended to replace the gasket every two years.
10.4 Cleaning the chamber
When working with steam, it is possible that traces of materials are de
-
posited on the walls of the chamber. It is recommended from time to time
to clean the inside of the chamber with a clean cloth. Also remove the
heater cover and clean the bottom of the chamber.